FAQ

The nominal sensitivity of the accelerometer defines the deviation specification within the acceptable sensitivity range.


For example: the nominal sensitivity of the accelerometer is defined as 100mV/g, then the sensitivity of the actual product is generally considered to be acceptable in the range of 90~110mV/g.

1. Connect the sensor's power supply line and voltage output line according to the sensor's instructions, and connect a multimeter to the output line.

2. Place the accelerometer sensor flat on a horizontal tabletop and fix it to the state of +1g as shown in the figure below. Ensure that the tabletop does not vibrate. Read the output voltage V1 (unit: mV) of the sensor on a multimeter; flip the sensor 180 degrees and place it. It is fixed on the desktop and is in the state of -1g as shown in the figure below. After it is stable, read the output voltage V2 (unit: mV) of the sensor on a multimeter. Then the sensitivity is calculated as: S=(V1-V2)/2 (unit: mV/g). Record the data and check whether the sensitivity on the calibration certificate is close.

Static Calibration

image.png

1、310V-1速度传感器信号RMS值算法的原理?

解答:310V-1速度传感器信号RMS值是由内部电子线路检测及自动换算的.

2. 310V-1速度传感器信号波动及响应时间较长.

解答:可以改进,并重新提供样品测试.

3. 310V-1速度传感器出线方式(能否做成侧端出线)

解答:可以做成侧端出线,外形见产品手册之型号340A

4. 310V-1速度传感器可否提供磁座(平底和马蹄形)

解答:目前只能提供平底的磁座

5.310V-1速度传感器改为电压信号输出(三线制)

解答:可以改为三线制电压信号输出,需要一定的数量或者做定制开发,需提供开发费.


关于以上问题如有不清楚,可以直接联系我司工程技术服务中心! 

技术服务热线:0755-85273639

Sample test can be provided, and the magnetic base needs to be matched when mass production


 可以定制我司的AVTS振动测试系统 (详见产品手册). 或者点击链接下载相应的资料 AVTS 振动测试标定系统 


The nominal sensitivity of the accelerometer defines the deviation specification within the acceptable sensitivity range.


For example: if the sensitivity of the accelerometer is defined as +-)

The nominal sensitivity of the accelerometer defines the deviation specification within the acceptable sensitivity range.


For example: if the sensitivity of the accelerometer is defined as +-)

The nominal sensitivity of the accelerometer defines the deviation specification within the acceptable sensitivity range.


For example: if the sensitivity of the accelerometer is defined as +-)